Remote Control - Command Description
R&S
®
UPP
738
User Manual 1411.1055.32 ─ 10
SENSe<1|2|3|6>:DATA:ALL?
<all>
Returns the measurement results of
all
channels in the form of float values which are
separated by a comma. Channels which are switched off are represented by value
"9.91E+37".
Dual channel instruments: 2 channels
Multi channel instruments: 1 to 8 channels
Suffix:
<1|2|3|6>
.
1 = function measurement results
2 = input or second monitor measurement results
3 = frequency , phase measurement results
6 = level monitor measurement results
Parameters:
<all>
float, float , float
Example:
INIT:CONT OFF; *WAI
SENS1:DATA:ALL?
Response:
1.0006 V,9.91E+37,1.00057 V,9.91E+37,1.00051
V,9.91E+37,9.91E+37,1.00051 V
Example:
INIT:CONT OFF; *WAI
SENS3:DATA:ALL?
Response:
1000 HZ,9.91E+37,-0.00248114
DEG,9.91E+37,0.00565384
DEG,9.91E+37,9.91E+37,0.0115933 DEG
A query with
MIN
or
MAX
returns all stored minimum and maxi-
mum measurement results.
Usage:
Query only
7.5.5
Output of Trace Data Sets
The commands are grouped for the following subsystems according to their function:
|
SWEepFFT
|
BARGraph
|
WAVEform
Suffix <n2> denotes the graphic window. The value range depends on the subsystem:
For the Subsystem Sweep:
<n2> = 1 … 4
For the Subsystem FFT:
<n2> = 1 | 2
For the Subsystem BARGraph:
<n2> = 1 | 2
For the Subsystems WAVeform:
<n2> = 1
●
DISPlay:BARGraph<n2>:SCANoffset
●
Triggering Measurements and Reading Results
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...