Instrument Functions
R&S
®
UPP
456
User Manual 1411.1055.32 ─ 10
For other level measurement options, see
Chapter 5.21, "Level Measurements"
Input Monitor
Used to activate the input level meter and to select the measurement method.
"Off"
The "Input Monitor " is switched off.
"Peak"
The peak input value is measured. The numerically highest input
sample is determined within a preset interval.
Note:
In contrast to peak measurement, which – like all measuring
functions – takes place downstream of the filters, no interpolation fil-
ter (oversampling filter) is used.
Remote command:
OFF | IPEak
5.36
Level Monitor
The level monitor is a level meter which is used to determine the level downstream of
the prefilter but upstream of the function-specific digital filters.
For other level measurement options, see
Chapter 5.21, "Level Measurements"
The level monitor works with period detection and measures the input signal – starting
at the zero crossing – across a complete period, but at least with 100
samples. With
function measurements with extremely short measuring times (e.g. peak measurement
with 1
ms interval time), the level monitor should therefore be switched off.
Level Monitor
Used to activate the level monitor and to select the measurement method.
"Off"
The level monitor is deactivated.
Level Monitor
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...