R&S SMA100A
Information on the Current Version and History
Release Notes for Firmware 1171.5360.52
8
Fixed Issues
● DME Modulation (R&S® SMA-K26): Changing the pulse repetition rate has no
effect, when DME mode is reply and squitter is on.
● In rare cases the level of the R&S® SMA100A is too high. Only if attenuator mode
is set to High Power Mode and ALC state is set to off.
1.5 Version 3.01.102.14
Released: October 2013
Modified Functionality
● Attenuator Settings: The lower limit of the level range is shifted around -5 dB (only
for R&S® SMA-B103 and R&S® SMA-B106)
Fixed Issues
● NRP-Z Power Viewer:
– After a factory preset the NRP-Z Power Viewer could not be switched on.
– In rare cases the NPR-Z Power Viewer in the block diagram is not readable.
● The dialog Software Options open up very slow.
● After a firmware update via LAN a message box appears.
● The display seems to be diffuse.
● The firmware update using ISO image does not create the directory /var/user.
1.6 Version 3.01.102.11
Released: October 2013
New Functionality
● Support of new synthesis extension board (R&S® SMA-B20, SYNEX) with part
number 1413.1800.06
● Support of new synthesis extension board (R&S® SMA-B22, SYNEX) with part
number 1413.1800.02
● Support of new rf output unit board with part number 1400.1707.07
● New remote control emulation for R&S® SME02, R&S® SME03, R&S® SME06
● New remote control emulation for Panasonic PA8303
● Remote update via LAN. For details see chapter 3.3
● Security Settings: