Remote Control Commands
R&S
®
RTM3000
699
User Manual 1335.9090.02 ─ 08
TRIGger:A:ARINc:LABel:CONDition
<Compare>
Define the operators for comparison of the decoded label condition with the specified
label.
Parameters:
<Compare>
EQUal | NEQual | GTHan | GEQual | LEQual | LTHan | WITHin |
OUTSide
INRange
: in range
OORange
: out of range
EQUal | NEQual | GTHan | LTHan
Equal, not equal, greater than, less than. These conditions
require one label to be set with
.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with
and
.
TRIGger:A:ARINc:LABel:MAXimum
<LabelMaximum>
Specifies the maximum value of the label if
TRIGger:A:ARINc:LABel:CONDition
is set to
INRange
or
OORange
.
Parameters:
<LabelMaximum>
01X-string
TRIGger:A:ARINc:LABel:MINimum
<LabelMinimum>
Sets the minimum value of the label to be triggerd on.
Parameters:
<LabelMinimum>
01X-string
TRIGger:A:ARINc:SDI
<SDIvalue>
Sets the source/destination identifier (SDI) bits to be triggered on.
Parameters:
<SDIvalue>
ANY | S00 | S01 | S10 | S11
TRIGger:A:ARINc:SSM
<SSMvalue>
Sets the sign/status matrix (SSM) bits to be triggered on.
Parameters:
<SSMvalue>
ANY | S00 | S01 | S10 | S11