Remote Control Commands
R&S
®
RTM3000
627
User Manual 1335.9090.02 ─ 08
*RST:
SBIT
TRIGger:A:UART:PATTern
<DataPattern>
Defines the bit pattern as trigger condition.
Parameters:
<DataPattern>
Binary pattern with max. 32 bit. Characters 0, 1, and X are
allowed.
*RST:
1 = "00000001"
TRIGger:A:UART:PLENgth
<PatternLength>
Defines how many symbols build up the serial pattern.
Parameters:
<PatternLength>
Number of symbols
Range:
1 to 4
Increment: 1
*RST:
1
TRIGger:A:UART:POFFset
<PatternByteOffset>
Sets the number of symbols before the first symbol of the pattern.
Parameters:
<PatternByteOffset> Number of ignored symbols
Range:
0 to 4095
Increment: 1
*RST:
0
16.11.4.3
UART - Decode Results
In all
BUS<b>:UART...
commands, the suffix <b> selects the bus.
.................................................................................................628
........................................................................................... 628
............................................................................................628
................................................................................. 628
BUS<b>:UART:RX:FRAMe<n>:STARt?
............................................................................628
BUS<b>:UART:TX:FRAMe<n>:STARt?
............................................................................ 628
..................................................................................629
............................................................................ 629
.............................................................................629
................................................................................. 629
BUS<b>:UART:RX:FRAMe<n>:STATe?
............................................................................629
BUS<b>:UART:TX:FRAMe<n>:STATe?
............................................................................ 629
.............................................................................. 629
BUS<b>:UART:RX:FRAMe<n>:WCOunt?
.........................................................................629
Serial Bus Analysis
www.allice.de
Allice Messtechnik GmbH