TRIGger:A:TYPE
<Type>
Sets the trigger typer.
Parameters:
<Type>
EDGE | WIDTh | TV | BUS | LOGic | LINE | RISetime | RUNT
EDGE
Edge trigger
WIDTh
Width trigger
TV
Video trigger
BUS
Requires at least one protocol option for serial bus (R&S
RTM-
K1 to K3)
LOGic
Pattern trigger, logic trigger
LINE
Trigger on power supply line
RISetime
Trigger on rise and fall time (transition trigger)
RUNT
Trigger on runts, on pulses with smaller amplitude
TRIGger:A:HOLDoff:MODE
<HoldOffMode>
Enables or disables the holdoff time.
Parameters:
<HoldOffMode>
TIME | OFF
*RST:
Off
TRIGger:A:HOLDoff:TIME
<HoldOffTime>
Defines the holdoff time. The next trigger occurs only after the holdoff time has passed.
Parameters:
<HoldOffTime>
Default unit: s
16.5.2
Edge Trigger
............................................................................................... 457
..........................................................................................457
........................................................................................... 457