Remote Control Commands
R&S
®
NGL200/NGM200
126
User Manual 1178.8736.02 ─ 09
OUTPut:TRIGgered
<arg0>
OUTPut:TRIGgered[:STATe]
<arg0>[, <Channel list>]
OUTPut:TRIGgered[:STATe]?
[<Channel list>]
Enables or disables the triggered event for output.
Setting parameters:
<arg0>
1
Trigger is enabled.
0
Trigger is disabled.
Parameters for setting and query:
<Channel list>
<list>
Example:
OUTP:TRIG 1
When a trigger event occurs, respective channel output is trig-
gered.
See
Example:
OUTPut:TRIG? (@1)
Queries trigger event for output at channel 1.
OUTPut:TRIGgered:BEHavior
<arg0>[, <Channel list>]
OUTPut:TRIGgered:BEHavior?
[<Channel list>]
Sets or queries output behavior when a trigger event occurs.
Setting parameters:
<arg0>
ON | OFF | GATed | <list>
ON
Output is set on when a trigger event occurs.
OFF
Output is set off when a trigger event occurs.
GATed
Output is set gated when a trigger event occurs.
Parameters for setting and query:
<Channel list>
<list>
Example:
OUTP:TRIG:STAT 1
OUTP:TRIG:BEH ON
Example:
OUTPut:TRIG:BEH? (@1)
Queries output behavior of trigger event at channel 1.
7.5.8
Range/DVM Setting
The DVM and range settings are available only with R&S NGM power supply series
equipped with R&S NGM-K104 (P/N: 3643.9927.02).
Configuration Commands