Instrument functions
R&S
®
LCX Series
73
User Manual 1179.2260.02 ─ 02
[Auto]
Selects the impedance range automatically, according to the connected sample com-
ponent.
Note:
If you have fixed the impedance range with
blocked.
Remote command:
[Hold]
Freezes the currently set range.
When activated, the R&S
LCX displays the prefix "Hold:" at the range value.
Remote command:
[Trig.]
Triggers one measurement cycle manually.
Remote command:
8.6
Measurement parameters
The settings described in this section refer to parameters that you can configure for
your specific measurement in addition to the test signal, measurement mode, or mea-
surement functions.
About measurement parameters
Aspects on the measurement parameters:
●
Measurement speed
The measurement speed defines the number of measurements in one second
when triggered continuously.
Impact of measurement speed and accuracy:
The measurement accuracy decreases with increasing speed, i.e. the higher the
number of measurements per second you perform, the lower the measurement
accuracy you can expect.
At low test signal frequencies, the R&S
LCX reduces the measurement speed
automatically.
See
●
Cable length
Any cable covers inductive and capacitive parts, which depend in particular on the
cable length. As these characteristics are considered as parasitics, they distort a
Measurement parameters
www.allice.de