
R&S FSP
Remote Control – Description of Commands
CALibration Subsystem
6.100
Operating Manual 1164.4556.12 - 06
6.6 CALibration Subsystem
The commands of the CALibration subsystem determine the data for system error correction in the
instrument.
CALibration:ABORt
This command aborts the acquisition of correction data and restores the last complete correction data
set.
Example:
"CAL:ABOR"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
CALibration[:ALL]?
This command initiates the acquisition of system error correction data. A "0" is returned if the
acquisition was successful.
In order to recognize when the acquisition of correction data is completed, the MAV bit in the status
byte can be used. If the associated bit is set in the Service Request Enable Register, the instrument
generates a service request after the acquisition of correction data has been completed.
Example:
"*CLS"
'
Resets the status management.
"*SRE 16"
'
Enables MAV bit in the Service Request Enable Register.
"*CAL?"
'Starts the correction data recording and then a service request is generated.
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
CALibration:RESult?
This command outputs the results of the correction data acquisition. The lines of the result table (see
chapter
“Recording the Correction Data – CAL Key” on page 4.49
) are
output as string data separated by commas:
Return value:
"Total Calibration Status: PASSED","Date (dd/mm/yyyy): 12/07/
2006",
"Time: 16:24:54","Runtime:00.06"
Example:
"CAL:RES?"
During the acquisition of correction data, the instrument does not accept any
remote control commands, except
*RST
CALibration:ABORt
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...