Operating Manual 1164.4556.12 - 06
4.49
R&S FSP
Instrument Functions
Analyzer Mode
TRACE
POSITION
The
TRACE POSITION
softkey activates the entry of the trace position for 0 differ-
ence. The position is stated in % of the diagram height.
The range of values extends from -100% to +200%
Remote command:
CALC:MATH:POS 50PCT
TRACE MATH
OFF
The
TRACE MATH OFF
softkey switches the math function off.
Remote command:
CALC:MATH:STAT OFF
4.5.9
Recording the Correction Data – CAL Key
The R&S FSP obtains its high measurement accuracy through its inbuilt self-align-
ment method.
The correction data and characteristics required for the alignment are determined by
comparison of the results at different settings with the known characteristics of the
high-precision calibration signal source of R&S FSP at 128 MHz. The correction
data are then available in the instrument as a file and can be displayed by means of
the
softkey.
For service purposes the use of correction data can be deactivated by means of the
CAL CORR ON/OFF
softkey. If the correction data recording is aborted, the last
complete correction data set is restored.
The
CAL
key opens a menu with the available functions for recording, displaying
and activating the data for self alignment.
CAL
CAL TOTAL
The
CAL TOTAL
softkey starts the recording of correction data of the instrument.
If the correction data recording has failed or if the correction values are deactivated
(
CAL CORR = OFF
softkey),
UNCAL
is displayed in the status field.
Remote command:
*CAL?
CAL ABORT
The
CAL ABORT
softkey interrupts the recording of correction data and restores the
last complete correction data set.
Remote command:
CAL:ABOR
The term "Calibration" formerly used for the integrated self alignment was often
mistaken for the "true" calibration of the instrument at the test set in production and
in service. It is therefore no longer used although it appears in the abbreviated
form in the name of keys ("CAL...").
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...
Page 796: ...R S FSP 4 14 Operating Manual 1164 4556 12 06 ...