Operating Manual 1164.4556.12 - 06
4.165
R&S FSP
Instrument Functions
Basic Settings
The following settings can be modified here:
• The
REFERENCE INT/EXT
softkey determines the source of the reference. For
details refer to section
“External Reference” on page 4.165
• The
NOISE SRC ON/OFF
softkey switches on and off the voltage supply for an
external noise source. For details refer to section
• The
PREAMP
softkey switches on the RF preamplifier gain. This softkey is only
available with option EL. ATTENUATOR (B25). For details refer to section
• The
TRANSDUCER
softkey opens a submenu for entering the correction
characteristics for transducers. For details refer to section
• The
GENERAL SETUP
softkey opens a submenu for all the general settings such
as GPIB address, date and time as well as the configuration of the device
interfaces.
FIRMWARE OPTIONS
can be installed under this menu item. For
details refer to section
• The
SYSTEM INFO
softkey opens a submenu for displaying the hardware
configuration of the instrument, the switching cycle statistics and system
messages. For details refer to section
• The
SERVICE
softkey opens a submenu in which special device functions and
system information can be selected for servicing. The password required for
service functions can be entered in this submenu. For details refer to section
.
• The
FIRMWARE UPDATE
opens a submenu to update the firmware version, to
restore the firmware and to update the firmware path. For details refer to section
“Firmware Update” on page 4.192
• The
SERVICE FUNCTIONS
softkey enables additional special settings for
servicing and troubleshooting. It is available after entering the corresponding
password under the
SERVICE
softkey.
4.6.3.1
External Reference
The R&S FSP can use the internal reference source or an external reference source
as frequency standard from which all internal oscillators are derived. A 10 MHz crys-
tal oscillator is used as internal reference source. In the default setting (internal ref-
erence), this frequency is available as output signal at rear-panel connector REF
OUT, e.g. to synchronize other instruments to the reference of the R&S FSP.
Side menu
!
!
Summary of Contents for R&S FSP Series
Page 1: ...R S FSP Spectrum Analyzer Operating Manual 1164 4556 12 06 Test Measurement Operating Manual ...
Page 24: ...R S FSP Putting into Operation 1 2 Operating Manual 1164 4556 12 06 ...
Page 72: ...R S FSP Manual Operation 3 2 Operating Manual 1164 4556 12 06 ...
Page 793: ...R S FSP Index Operating Manual 1164 4556 12 06 4 11 ...
Page 794: ...R S FSP 4 12 Operating Manual 1164 4556 12 06 ...
Page 795: ...Operating Manual 1164 4556 12 06 4 13 R S FSP ...