Remote Commands
R&S
®
Spectrum Rider FPH
352
User Manual 1321.1011.02 ─ 06.00
This command is a query and therefore has no *RST value.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<Return values>
returns one value for every harmonic.
The first value is the absolute power of the first harmonic. The
levelunit depends on the unit you have set with the
on page 297 command. The other val-
ues are power levels relative to the first harmonic. The unit for
these is dB.
The total number of return values depends on the number of
MARKer<1...6>:FUNCtion:HARMonics:NHARmonics
Example:
See
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:NHARmonics
<NofHarmonics>
This command sets the number of harmonics. The valid range is from 1 to 6 harmon-
ics.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<NofHarmonics>
Range:
1 to 6
*RST:
2
Example:
See
CALCulate<1...2>:MARKer<1...6>:FUNCtion:
CALCulate<1...2>:MARKer<1...6>:FUNCtion:HARMonics:PRESet
This command optimizes the settings for the harmonic distortion measurement.
This command is an event and therefore has no query and no *RST value.
The suffix at CALCulate and MARKer are irrelevant for this command.
Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
CALC:MARK:FUNC:HARM:NHAR 4
Sets the number of harmonics to 4.
CALC:MARK:FUNC:HARM:PRES
Adjusts the settings of the measurement.
CALC:MARK:FUNC:HARM:LIST?
CALC:MARK:FUNC:HARM:DIST? TOT
Queries the position of the harmonics and the total harmonic dis-
tortion.
Remote Control - Commands