Remote Commands
R&S
®
Spectrum Rider FPH
346
User Manual 1321.1011.02 ─ 06.00
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel[:RELative]
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel[:RELative]:
STATe
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel:ABSolute
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel:ABSolute[:
STATe]
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel:RESult?
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ALTernate<1...11>[:
RELative]
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ALTernate<1...11>[:
RELative]:STATe
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ALTernate<1...11>:
ABSolute
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ALTernate<1...11>:
ABSolute[:STATe]
●
CALCulate<1...2>:LIMit<1...2>:ACPower:ALTernate<1...11>:
RESult?
CALCulate<1...2>:LIMit<1...2>:ACPower:STATe
<State>
This command turns the limit check for ACLR measurements on and off.
You have to use
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel[:
CALCulate<1...2>:LIMit<1...2>:
ACPower:ALTernate<1...11>[:RELative]:STATe
with this command to select the channels the limit check should be performed on.
The suffix at CALCulate and LIMit are irrelevant for this command.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:LIM:ACP ON
Activates the limit check.
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel[:RELative]
<RelativeLimit>
This command defines the relative limit of the adjacent channels. The reference value
for the relative limit value is the measured channel power.
Note that the relative limit has no effect on the limit check if it is below the absolute limit
CALCulate<1...2>:LIMit<1...2>:ACPower:ACHannel:ABSolute
on page 347. This mechanism allows automatic checking of the absolute basic values
of adjacent-channel power as defined in mobile radio standards.
Remote Control - Commands