Spectrum Analyzer Mode
R&S
®
Spectrum Rider FPH
177
User Manual 1321.1011.02 ─ 06.00
Figure 6-11: Gate trigger measurement
1 = Gate trigger settings mode
2 = Gate delay
3 = Gate length
4 = RF frequency
5 = Gate trigger softkeys menu
6 = Span 0 Hz (time domain)
7 = Measurement of LTE signal within the gate length
6.2.6
Working with Traces
The trace menu contains all functions available to customize the trace display.
●
..................................................................................... 177
...........................................................................................178
............................................................................... 180
................................................................................181
..................................................................................... 182
6.2.6.1
Selecting the Trace Mode
The R&S Spectrum Rider provides several trace modes. The trace mode defines the
way the R&S Spectrum Rider writes the trace.
1. Press the [TRACE] key.
2. Select the "Trace Mode" softkey to set the trace mode.
3. Select the trace mode you want to work with.
If you have selected the average trace mode ("Average: 10" menu item), the R&S
Spectrum Rider opens an input field to set the number of sweeps the R&S Spec-
trum Rider includes in the averaging.
4. Enter the number of sweeps to include in the averaging.
In continuous sweep mode, the R&S Spectrum Rider now calculates the moving
average over the number of sweeps you have specified. In single sweep mode, it
stops the measurement after finishing the sweeps and averages the traces.
The R&S Spectrum Rider provides the following trace modes.
●
Clear/Write
Configuring Spectrum Measurements