Setting up a measurement
R&S
®
NRPM
46
Mannual 1425.8663.02 ─ 08
To connect RF frontend with the sensor module in an RF test chamber
1.
NOTICE!
Risk of antenna damage. Observe
on page 18.
In the RF test chamber, mount the antenna modules.
For information on the connecting options, refer to the documentation of the
antenna equipment.
2. On the interface module
using a torx screwdriver TX8.
3.
NOTICE!
Risk of broken cables. The connectivity of the antenna cable ends is
sensitive, and therefore the contact durability of the connection is limited.
Do not connect and disconnect the cables to the interface module frequently.
Connect each antenna module cable to the antenna module cable connectors
of the interface module.
4. Mount the strain relieve of the interface module.
5. To connect the R&S
NRPM-Z3 interface module to the R&S
NRPM3(N) sen-
R&S NRPM-Z3 connected to R&S NRPM3
a) Make sure that the sensor module is powered off, i.e. not yet connected to
the power sourcing interface, see
b) Insert the 16-pin sensor connector of the
straight into the antenna connector of the R&S
NRPM3(N). Take care that
the guide gap of the sensor connector fits to the guide lug of the mating
antenna connector of the sensor module.
6. Continue with
Connecting to the controller PC
To disconnect the R&S
NRPM-Z3 from the sensor module
1. Disconnect the power sourcing interface of the sensor module.
2. Pull the sensor connector of the interface module from the antenna connector
of the sensor module. Take care not to tilt it.
6.3.2
Connecting to the controller PC
The controlling host of the R&S
NRPM OTA power measurement solution is a
computer, using a supported software for controlling the R&S
NRPM3(N) sensor
modules, see
Chapter 4.5, "Hardware and software requirements"
Connecting an R&S
NRPM OTA power measurement