Troubleshooting
R&S
®
NRPM
205
Mannual 1425.8663.02 ─ 08
Firmware update was aborted
If there is not enough free memory space, the firmware update aborts. An error
message is displayed, and the status LED of the sensor module starts flashing
red.
1. Execute the sanitzation procedure, as described in the instrument security
procedures. You find this document on the product page at:
.
2. Start the firmware update again.
12.4
Cannot establish a LAN connection
If you have problems to establish a LAN connection as described in
ter 6.3.4, "Using the LAN connection"
on page 50, try the following measures:
●
Use the
Configure Network Sensor
component of the R&S
NRP
Toolkit, see
Chapter 12.4.1, "Components of the R&S
●
"Troubleshooting for peer-to-peer connections"
12.4.1
Components of the R&S
NRP
Toolkit
Access:
► "Start"
> "NRP-Toolkit"
The following tools are part of the R&S
NRP
Toolkit for Windows.
Configure network sensor
Useful if you have troubles establishing a LAN connection with an R&S
NRPM3N
LAN sensor module. The tool provides the following functions:
●
Configuring the network settings by (temporarily) connecting the selected sen-
sor to the computer using USB.
●
Discovering the sensors that have been configured via the Zeroconf (APIA)
protocol.
The tool comes with a guide (PDF) that is also available in the "Start" menu. The
guide explains the network setup.
Cannot establish a LAN connection