Remote control commands
R&S
®
NRPM
166
Mannual 1425.8663.02 ─ 08
......................................................................166
.......................................................................166
[SENSe<Sensor>:]AVERage:TCONtrol
.................................................................. 167
[SENSe<Sensor>:]AVERage[:STATe]
..................................................................... 167
[SENSe<Sensor>:]AVERage:COUNt
<count>
Sets the number of readings that are averaged for one measured value. The
higher the count, the lower the noise, and the longer it takes to obtain a measured
value.
Average count is often also called averaging factor, but it designates the same
parameter, i.e the number of measured values that have to be averaged for form-
ing the measurement result.
Averaging is only effective, when
[SENSe<Sensor>:]AVERage[:STATe]
is
turned on.
Parameters:
<count>
Range:
1 to 65536
*RST:
4
Example:
AVERage:COUNt 4
Manual operation:
[SENSe<Sensor>:]AVERage:RESet
Deletes all previous measurement results that the averaging filter contains and
initializes the averaging filter. The filter length gradually increases from 1 to the
set averaging factor. Thus, trends in the measurement result become quickly
apparent. Note that the measurement time required for the averaging filter to set-
tle completely remains unchanged.
Use this command if:
●
A high averaging factor is set.
●
Intermediate values are output as measurement results.
[SENSe<Sensor>:]AVERage:TCONtrol MOVing
●
Power has significantly decreased since the previous measurement, for exam-
ple by several powers of 10.
Configuring basic measurement parameters