Remote control commands
R&S
®
NRPM
142
Mannual 1425.8663.02 ─ 08
Moving termination control
Returns intermediate values to facilitate early detection of changes in the mea-
sured quantity. For each partial measurement, the sensor module returns a new
average value as an intermediate measurement result. Thus, the individual result
is a moving average of the last partial measurements. The parameter average
count defines how many partial measurements are included.
Select the moving termination control when you want to observe and detect
trends in the results during an ongoing measurement.
10.5.3
Interplay of the controlling mechanisms
The following examples use continuous measurement scenarios. But these sce-
narios apply also to single measurements. The only difference is that a single
measurement is not repeated.
10.5.3.1
Continuous average mode
General settings for these examples:
●
●
●
[SENSe<Sensor>:]AVERage[:STATe]
Controlling the measurement