Remote Control - Commands
NRP2
196
User Manual 1173.9140.02 ─ 07
*ESR?
<value>
E
vent
S
tatus
R
ead returns the contents of the event status register in decimal form (0
to 255) and subsequently sets the register to zero.
Parameters:
<Value>
0 to 255
Usage:
Query only
*IDN?
ID
e
N
tification queries the instrument identification.
Note:
The return string can be customized using
SYSTem:IDN
and
SYSTem:IDN:AUTO
commands
Example:
Return values
Rohde&Schwarz, NRP2, 10001, 07.03
The strings have the following meanuing:
Rohde&Schwarz
= manufacturer
NRP2
= instrument type
10001
= serial number
07.03
= firmware version
Usage:
Query only
*IST?
<status_bit>
I
ndividual
ST
atus query returns the current value of the IST flag in decimal form. The
IST flag is the status bit which is sent during a parallel poll (
and Parallel Poll Enable Register (PPE)"
Parameters:
<StatusBit>
0 | 1
Usage:
Query only
*OPC
Setting Command :
*OPC
OP
eration
C
omplete sets bit 0 in the event status register after all preceding com-
mands have been executed. This bit can be used to initiate a service request.
*OPC
must be sent at the end of a program message.
Query:
*OPC?
OP
peration
C
omplete query returns a
1
when all previous commands have been pro-
cessed. It is important that the timeout set on the IEC/IEEE bus is sufficiently long.
Common Commands to IEEE 488.2