Remote Control - Commands
NRP2
305
User Manual 1173.9140.02 ─ 07
Bit No.
Meaning
6
MSS: Master-Status summary bit
This bit is set if the device triggers a service request. This is the case if one of the other bits
of this register is set together with its enable bit in the Service Request Enable register
SRE.
7
Operation Status Register summary bit
This bit is set if an EVENt bit is set in the Operation Status Register and the associated
ENABLe bit is set to 1. A set bit denotes that an action is being performed by the device.
Information on the type of action can be obtained by querying the Operation Status Regis-
ter.
6.13.6 IST Flag and Parallel Poll Enable Register (PPE)
Similar to the SRQ, the IST flag combines the complete status information in a single
bit. It can be queried by a parallel poll (see
chapter 6.13.20.3, "Parallel Poll"
*IST?
command.
The Parallel Poll Enable Register (PPE) determines which bits of the STB affect the
IST flag. The bits of the STB are ANDed with the corresponding bits of the PPE; bit 6 is
also used - in contrast to the SRE. The IST flag is obtained by ORing all results
together.
The PPE can be set by the
*PRE
command and read by the
*PRE?
query.
6.13.7 Device Status Register
This register contains information on current device states (CONDition register) or
states that occurred since the last query (EVENt register).
The register can be read by the queries
STATus:DEVice:CONDition?
or
STATus:DEVice[:EVENt]?
.
Table 6-19: Meaning of bits used in the Device Status Register
Bit No.
Meaning
0
Not used
1
Sensor A connected
2
Sensor B connected
3
Sensor A error
Sensor A is erroneous.
4
Sensor B error
Sensor B is erroneous.
5
Sensor A Front/Rear
Indicates whether sensor A is connected at the front (bit is not set) or rear (bit is set) of the
device.