Browser-Based User Interface
R&S
®
NRPxxA(N)
50
User Manual 1177.6017.02 ─ 05
6.5.2
Averaging Settings
Describes the parameters for automatic averaging.
Further information:
●
Chapter 9.8.1, "Configuring Auto Averaging"
Access: main dialog of the web user interface > navigation pane > "Averaging"
............................................................................................... 51
Averaging Mode
Groups the averaging settings. See also
Chapter 9.8.1, "Configuring Auto Averaging"
<Mode> ← Averaging Mode
Sets the averaging mode.
"manual"
Disables auto averaging. Enter the average count under
"auto 1
dB" / "auto 0.1
dB" / "auto 0.01
dB" / "auto 0.001
dB"
Uses an automatic averaging filter with the respective resolution
index.
"noise content" Predefines the compliance to an exactly defined noise component.
Enter this value under
Remote command:
[SENSe<Sensor>:]AVERage[:STATe]
[SENSe<Sensor>:]AVERage:COUNt:AUTO
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE
<Value> ← Averaging Mode
The content of this field depends on the setting under
●
If "manual" is set:
Sets the average count, also called averaging factor.
●
If "auto xx
dB" is set:
Displays the resolution index.
●
If "noise content" is set:
Settings