Instrument functions
R&S
®
NGP800
99
User Manual 5601.5610.02 ─ 09
6.17.2
Channel adjustment
The "Adjustment" calculates the required adjustment coefficient internally for voltage
and current on the selected channel.
Channel adjustment setup
Recommended instruments
●
Digital multimeter (DMM): 6
½
digits
●
Shunt resistor: 10 m
Ω
, at least 25 A rating, 0.02 % accuracy,
●
Connect the DMM to the instrument and monitor the measured voltage or current.
See
and
.
Figure 6-71: Voltage adjustment setup
Adjustment