SCPI Commands HMF series
Remote Control
66
Command Reference
SCPI Programmers Manual
2.7 Burst commands
BURSt:NCYCles {<Number of cycles> | MINimum | MAXimum}
............................................. 66
BURSt:NCYCles? [MINimum | MAXimum]
................................................................................ 67
BURSt:INTernal:PERiod {<Period> | MINimum | MAXimum}
.................................................... 67
BURSt:INTernal:PERiod? [MINimum | MAXimum]
.................................................................... 67
BURSt:PHASe {<Phase> |MINimum | MAXimum}
.................................................................... 67
BURSt:PHASe? [MINimum | MAXimum]
................................................................................... 67
BURSt:STATe {OFF | ON | 0 | 1}
BURSt:MODE <Mode>
Defines the burst mode.
Parameters:
<Mode>
TRIGgered | GATed
TRIGgered
:
A trigger will generate a burst with a predefined
number of cycles.
GATed
:
The signal will be either on or off, depending on the
level of the external signal at the „Trigger input/ouput“
connector. If the gate signal is „true“ (high +5 V), the
function generator will deliver a continuous signal
until the „gate“ closes (0V TTL low). If no power is ap-
plied to the TRIG INPUT connector, the output signal
will stop as the function generator will stop generati-
on.
BURSt:MODE?
Queries the burst mode.
Return values:
TRIG | GAT
BURSt:NCYCles {<Number of cycles> | MINimum | MAXimum}
Defines the number of burst cycles.
Parameters:
<Number of cycles>
Numeric value
MIN
: 1.0E+00
MAX
: 5.04
*RST:
1.0E+00