
SCPI Commands HMF series
Remote Control
64
Command Reference
SCPI Programmers Manual
SWEep:TIME? [MINimum | MAXimum]
Queries the sweep time.
Return values:
Numeric value in sec.
MIN
: 1.000000E-03
MAX
: 5.02
SWEep:STATe {OFF | ON | 0 | 1}
Activates or deactivates the sweep function.
Parameters:
<State>
ON | 1
:
The sweep function will be activated.
OFF | 0
:
The sweep function will be disabled.
*RST: OFF | 0
SWEep:STATe?
Queries the state of the sweep function.
Return values:
ON
:
The sweep function is activated.
OFF
:
The sweep function is disabled.
MARKer:FREQuency {<Frequency> | MINimum | MAXimum}
Defines the sweep marker frequency. The marker frequency have be set between the start and
stop frequency. If the signal frequency reaches the marker frequency a signal will be generated
available at the TRIG OUTPUT connector.
Parameters:
<Time>
Numeric value in Hz depending on the start and stop frquency.
MARKER:FREQuency? [MINimum | MAXimum]
Queries the sweep marker frequency.
Return values
:
Numeric value in Hz depending on the start and stop frquency.
MARKer {OFF | ON | 0 | 1}
Activates or deactivates the sweep marker function.
Parameters:
<State>
ON | 1
:
The sweep marker function will be activated.
OFF | 0
:
The sweep marker function will be disabled.
*RST: OFF | 0