Remote Commands
R&S
®
FSW
953
User Manual 1173.9411.02 ─ 19
Example:
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all
delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Manual operation:
See
11.8.3.8
Marker Peak Lists
Useful commands for peak lists described elsewhere
●
●
●
chapter 11.8.3.3, "Configuring and Performing a Marker Search"
Remote commands exclusive to peak lists
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
CALCulate:MARKer:FUNCtion:FPEaks:COUNt?
............................................................... 953
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks[:IMMediate]
.............................................. 954
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE
................................................ 954
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT
...................................................... 954
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT
.......................................................955
CALCulate:MARKer:FUNCtion:FPEeaks:X?
..................................................................... 955
CALCulate:MARKer:FUNCtion:FPEeaks:Y?
..................................................................... 955
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
<State>
This command turns labels for peaks found during a peak search on and off.
The labels correspond to the marker number in the marker peak list.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
CALC:MARK:FUNC:FPE:ANN:LAB:STAT OFF
Removes the peak labels from the diagram
Manual operation:
See
CALCulate:MARKer:FUNCtion:FPEaks:COUNt?
This command queries the number of peaks that have been found during a peak
search.
Analyzing Measurements (Basics)