![Rohde & Schwarz FSW13 User Manual Download Page 906](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847906.webp)
Remote Commands
R&S
®
FSW
906
User Manual 1173.9411.02 ─ 19
Return values:
<SideBand>
NORMal
The sideband at the output is identical to the RF signal.
INVerted
The sideband at the output is the inverted RF signal sideband.
Example:
OUTP:IF IF2
Activates output at the IF OUTPUT (2 GHZ) connector.
OUTP:IF:SBAN?
Queries the sideband provided at the connector.
Usage:
Query only
Manual operation:
See
OUTPut:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FSW.
Parameters:
<State>
ON
User port is switched to OUTPut
OFF
User port is switched to INPut
*RST:
OFF
OUTPut:UPORt[:VALue]
<Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit
7
6
5
4
3
2
1
0
Pin
N/A
N/A
5
3
4
7
6
2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern. If the user port is programmed
to input instead of output (see
on page 867), the output value is
temporarily stored.
Parameters:
<Value>
bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range:
#B00000000 to #B00111111
Example:
OUTP:UPOR #B00100100
Sets pins 5 and 7 to 5 V.