Remote Commands
R&S
®
FSW
834
User Manual 1173.9411.02 ─ 19
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Usage:
Event
SCPI confirmed
Manual operation:
See
See
[SENSe:]FREQuency:STARt
<Frequency>
This command defines a start frequency for measurements in the frequency domain.
Parameters:
<Frequency>
0 to (fmax - min span)
*RST:
0
Example:
FREQ:STAR 20MHz
Usage:
SCPI confirmed
Manual operation:
See
See
[SENSe:]FREQuency:STOP
<Frequency>
This command defines a stop frequency for measurements in the frequency domain.
Parameters:
<Frequency>
min span to fmax
*RST:
fmax
Example:
FREQ:STOP 2000 MHz
Usage:
SCPI confirmed
Manual operation:
See
See
11.7.1.2
Configuring Signal Tracking
When signal tracking is activated, the maximum signal is determined after each fre-
quency sweep and the center frequency is set to the frequency of this signal. Thus with
drifting signals the center frequency follows the signal.
For more details see
chapter 6.3.1, "Impact of the Frequency and Span Settings"
CALCulate:MARKer:FUNCtion:STRack[:STATe]
............................................................... 835
CALCulate:MARKer:FUNCtion:STRack:BANDwidth
.......................................................... 835
CALCulate:MARKer:FUNCtion:STRack:THReshold
...........................................................835
CALCulate:MARKer:FUNCtion:STRack:TRACe
................................................................ 835
Setting Basic Measurement Parameters