Remote Commands
R&S
®
FSW
835
User Manual 1173.9411.02 ─ 19
CALCulate:MARKer:FUNCtion:STRack[:STATe]
<State>
This command turns signal tracking on and off.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
CALCulate:MARKer:FUNCtion:STRack:BANDwidth
<Bandwidth>
This command defines the bandwidth around the center frequency that is included in
the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Parameters:
<Bandwidth>
Range:
10 Hz to Max span
*RST:
(= span/10 on activating the function)
Default unit: Hz
Manual operation:
See
CALCulate:MARKer:FUNCtion:STRack:THReshold
<Level>
This command defines the threshold level for the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Parameters:
<Level>
The unit depends on
Range:
-130 dBm to 30 dBm
*RST:
-120 dBm
Manual operation:
See
CALCulate:MARKer:FUNCtion:STRack:TRACe
<TraceNumber>
This command selects the trace on which the largest signal is searched for.
Parameters:
<TraceNumber>
1 to 6
Range:
1 to 6
*RST:
1
Manual operation:
See
Setting Basic Measurement Parameters