Remote Commands
R&S
®
FSW
814
User Manual 1173.9411.02 ─ 19
<ElAttenuation>
Defines the electronic attenuation for a list entry.
Range:
0 to 30
Increment: 1
Default unit: dB
<FilterType>
Selects the filter type for a list entry. For more information see
<RBW>
Defines the resolution bandwidth for a list entry.
<VBW>
Defines the video bandwidth for a list entry.
<MeasTime>
Defines the measurement time for a list entry.
Range:
1 µs to 16000 s
<TriggerLevel>
The trigger level must be 0.
Return values:
<PowerLevel>
Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with
The result is a list of floating point values separated by commas.
The unit depends on
[SENSe:]LIST:POWer:SET
<PeakPower>, <RMSPower>, <AVGPower>,
<TriggerSource>, <TriggerSlope>, <TriggerOffset>, <GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level
after
sending this command.
Parameters:
<PeakPower>
ON | OFF | 0 | 1
Turns peak power evaluation on and off.
*RST:
1
<RMSPower>
ON | OFF | 0 | 1
Turns RMS power evaluation on and off.
*RST:
0
<AVGPower>
ON | OFF | 0 | 1
Turns average power evaluation on and off.
*RST:
0
Configuring and Performing Measurements