Remote Commands
R&S
®
FSW
801
User Manual 1173.9411.02 ─ 19
CALC:MARK:FUNC:TOI:SEAR ONCE
//Initiates a search for signals in the current trace.
//---------------Retrieving Results-------------
CALC:MARK:FUNC:TOI:RES?
//Returns the TOI.
11.5.12 Measuring the AM Modulation Depth
All remote control commands specific to AM modulation depth measurements are
described here.
Configuring and Performing the Measurement
..................................................... 801
Example: Measuring the AM Modulation Depth
....................................................802
11.5.12.1
Configuring and Performing the Measurement
The following commands control the measurement.
Useful commands for AM modulation depth described elsewhere
●
●
●
Remote commands exclusive to AM modulation depth measurements
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]
...................................................801
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:SEARchsignal ONCE
............................... 801
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:RESult?
...................................................802
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]
<State>
This command turns the AM Modulation Depth measurement on and off.
To work correctly, the measurement requires an AM modulated signal.
Parameters:
<State>
ON | OFF
*RST:
OFF
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth:SEARchsignal ONCE
This command initiates a search for the signals required for the AM depth measure-
ment.
Note that the command does not perform a new measurement, but looks for the sig-
nals on the current trace.