
Remote Commands
R&S
®
FSW
796
User Manual 1173.9411.02 ─ 19
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics[:STATe]
<State>
This command turns the harmonic distortion measurement on and off.
Note the following:
●
If you perform the measurement in the frequency domain, the search range for the
frequency of the first harmonic, whose power is determined, is defined by the last
span.
●
If you perform the measurement in the time domain, the current center frequency is
used as the frequency of the first harmonic. Thus, the frequency search is
bypassed. The first harmonic frequency is set by a specific center frequency in
zero span before the harmonic measurement is started.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
Manual operation:
See
11.5.10.2
Configuring the Measurement
The following commands control the harmonic distortion measurement.
Useful commands for harmonic distortion measurements described elsewhere
●
●
Remote commands exclusive to harmonic distortion measurements
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
...............................796
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics
......................................797
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:PRESet
..............................................797
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
<State>
This command selects the resolution bandwidth of the harmonic in respect to the band-
width of the first harmonic.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
identical
ON | 1
a multiple
*RST:
1
Manual operation:
See
Configuring and Performing Measurements