Remote commands exclusive to spurious emission measurements
CALCulate<n>:PSEarch|PEAKsearch:AUTO
.................................................................... 774
CALCulate<n>:PSEarch|PEAKsearch:MARGin
................................................................. 774
CALCulate<n>:PSEarch|PEAKsearch:PSHow
.................................................................. 774
CALCulate<n>:PSEarch|PEAKsearch:SUBRanges
........................................................... 774
CALCulate<n>:PSEarch|PEAKsearch:AUTO
<State>
This command turns the list evaluation on and off.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
CALC:PSE:AUTO OFF
Deactivates the list evaluation.
CALCulate<n>:PSEarch|PEAKsearch:MARGin
<Threshold>
This command defines the threshold of the list evaluation.
Parameters:
<Margin>
Range:
-200 to 200
*RST:
200
Default unit: dB
Example:
CALC:PSE:MARG 100
Sets the threshold to 100 dB.
CALCulate<n>:PSEarch|PEAKsearch:PSHow
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:PSE:PSH ON
Marks all peaks with blue squares.
CALCulate<n>:PSEarch|PEAKsearch:SUBRanges
<NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S
FSW stops the peak search and
continues the search in the next measurement range.
Parameters:
<NumberPeaks>
Range:
1 to 50
*RST:
25