Remote Commands
R&S
®
FSW
747
User Manual 1173.9411.02 ─ 19
[SENSe:]ESPectrum<sb>:RANGe<range>:SWEep:TIME:AUTO
<State>
This command turns automatic selection of the sweep time for a SEM range on and
off.
In case of high speed measurements, the sweep time has to be identical for all ranges.
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range>
1..30
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
ESP:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<range>:TRANsducer
<Transducer>
This command selects a transducer factor for a SEM range.
Note that
●
the transducer must cover at least the span of the range
●
the x-axis has to be linear
●
the unit has to be dB
Suffix:
<sb>
.
1 | 2 | 3
Sub block in a Multi-SEM measurement
<range>
1..30
Selects the measurement range.
Parameters:
<Transducer>
String containing the transducer file name, including the path
information.
Manual operation:
See
11.5.6.5
Configuring the Reference Range
The following commands define the reference range for the SEM sweep list.
...................................................................................... 748
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa
................................................................ 748
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe]
.............................................................. 748
................................................................................748
.................................................................................... 749
Configuring and Performing Measurements