Common Measurement Settings
R&S
®
FSW
415
User Manual 1173.9411.02 ─ 19
If the Sequencer is off, only the evaluation for the currently displayed measurement
channel is updated.
For details on the Sequencer, see
chapter 4.5.1, "The Sequencer Concept"
Remote command:
Continuous Sweep/RUN CONT
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
While the measurement is running, the "Continuous Sweep" softkey and the RUN
CONT key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note:
Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only
controls the sweep mode for the currently selected channel; however, the sweep mode
only has an effect the next time the Sequencer activates that channel, and only for a
channel-defined sequence. In this case, a channel in continuous sweep mode is swept
repeatedly.
Furthermore, the RUN CONT key controls the Sequencer, not individual sweeps. RUN
CONT starts the Sequencer in continuous mode.
For details on the Sequencer, see
chapter 4.5.1, "The Sequencer Concept"
Remote command:
Continue Single Sweep
After triggering, repeats the number of sweeps set in "Sweep Count", without deleting
the trace of the last measurement.
While the measurement is running, the "Continue Single Sweep" softkey and the RUN
SINGLE key are highlighted. The running measurement can be aborted by selecting
the highlighted softkey or key again.
Remote command:
Spectrogram Frames
These settings are only available if spectrogram display is active (see
"How to Display and Configure a Spectrogram"
Selecting a frame to display ← Spectrogram Frames
Selects a specific frame, loads the corresponding trace from the memory, and displays
it in the Spectrum window.
Note that activating a marker or changing the position of the active marker automati-
cally selects the frame that belongs to that marker.
Bandwidth, Filter and Sweep Configuration