Common Measurement Settings
R&S
®
FSW
370
User Manual 1173.9411.02 ─ 19
"Table"
Defines the conversion loss via the table selected from the list. Pre-
defined conversion loss tables are often provided with the external
mixer and can be imported to the R&S
FSW. Alternatively, you can
define your own conversion loss tables. Imported tables are checked
for compatibility with the current settings before being assigned.
Conversion loss tables are configured and managed in the
For details on conversion loss tables, see
on page 361.
For details on importing tables, see
Remote command:
Average for range 1:
Table for range 1:
Average for range 2:
Table for range 2:
Basic Settings
The basic settings concern general use of an external mixer. They are only available if
the
is "On".
...........................................................................372
LO Level
Defines the LO level of the external mixer's LO port. Possible values are from 13.0
dBm to 17.0 dBm in 0.1 dB steps. Default value is 15.5 dB.
Remote command:
Data Input and Output