Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1047
User Manual 1178.8520.02 ─ 08
Manual operation:
See
Commands for test, alignment, and diagnosis
The following commands are required to test and optimize the connection after it has
initially been set up.
[SENSe:]EFRontend<fe>:ALIGnment<ch>:FILE
..............................................................1047
[SENSe:]EFRontend<fe>:ALIGnment<ch>:STATe
........................................................... 1047
[SENSe:]EFRontend<fe>:FWUPdate
............................................................................. 1048
[SENSe:]EFRontend<fe>:SELFtest?
.............................................................................. 1048
[SENSe:]EFRontend<fe>:SELFtest:RESult?
................................................................... 1048
[SENSe:]EFRontend<fe>:ALIGnment<ch>:FILE
<File>
Selects or queries the touchstone file that contains correction data to compensate for
signal losses in the cable occurring at different IF signal frequencies.
Suffix:
<fe>
.
1
Connected frontend
<ch>
1..n
Currently irrelevant
Parameters:
<File>
string in double quotes
Path and file name of the correction data file. The file must be in
s2p
format.
If the specified file is not found or does not have the correct for-
mat, an error message is returned (
-256, "File name not
found"
,
-150, "String data error"
).
Example:
EFR:ALIG:FILE "FE44S.s2p"
Manual operation:
See
"Load Cable Correction S2p File"
[SENSe:]EFRontend<fe>:ALIGnment<ch>:STATe
<State>
Activates correction of the IF signal due to cable loss from the frontend to the analyzer.
Specify the file with correction data using
Suffix:
<fe>
.
1
Connected frontend
<ch>
1..n
Currently irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
Setting basic measurement parameters