![Rohde & Schwarz FSUP8 Manual Download Page 2](http://html.mh-extra.com/html/rohde-and-schwarz/fsup8/fsup8_manual_1477841002.webp)
Contents
R&S
®
FSUP
2
Instrument Security Procedures 1171.5702.52 ─ 02
Contents
1 Overview................................................................................................. 2
2 Instrument Models Covered..................................................................2
3 Security Terms and Definitions............................................................ 3
4 Types of Memory and Information Storage in the R&S
FSUP........... 3
5 Instrument Declassification.................................................................. 5
6 Special Considerations for USB Ports.................................................7
1 Overview
In many cases, it is imperative that the R&S
FSUP Signal Source Analyzers are used
in a secured environment. Generally these highly secured environments do not allow
any test equipment to leave the area unless it can be proven that no user information
leaves with the test equipment. Security concerns can arise when devices need to
leave a secured area e.g. to be calibrated or serviced.
This document describes the types of memory and their usage in the R&S
FSUP. It
provides a statement regarding the volatility of all memory types and specifies the
steps required to declassify an instrument through memory clearing or sanitization pro-
cedures. These sanitization procedures are designed for customers who need to meet
the requirements specified by the US Defense Security Service (DSS).
2 Instrument Models Covered
Table 2-1: Signal Source Analyzer models
Product name
Order number
R&S
FSUP8
1166.3505.09
R&S
FSUP26
1166.3505.27
R&S
FSUP50
1166.3505.51