5.9
R&S FSQ
Measuring Harmonics of Sinusoidal Signals
➢
Press the
MARKER 3
softkey in the marker menu.
Marker 3 is activated as the delta marker (Delta 3 [T1]). It automatically appears on the
next largest harmonic of the signal. The frequency offset and level offset from marker 1
on the fundamental are displayed in the marker field at the top edge of the screen. (see
Fig. 5-2
Measuring the harmonic suppression of the internal reference generator.
Markers delta 2 [T1] and delta 3 [T1] show the offset of the first and second
harmonics from the fundamental.