R&S
FS-K30
General Information
Quick
Start Guide
Software Manual 1300.6637.42 - 03
15
Fig. 1 Preparation for calibration
1. Set the
Second Stage Correction
parameter in the Measurement Setting view to
ON. Calibration cannot be performed until this parameter is set.
2. Start the calibration of R&S FS-K30 by pressing the
CAL
hotkey.
During calibration, the text "
Running ...
" is displayed in the Status Bar at the bottom
of the screen. The progress bar indicates the progress through the calibration
measurement.
After successful calibration, the Status Bar will display “
Measurement Complete
”
and the title bar at the top of the screen will show a
CALIBRATED
status on the
right-hand-side.
Note that this calibration measurement calibrates only the R&S FS-K30 application
and not the spectrum analyzer itself.
1.5.2 Performing the amplifier measurement
After calibration has been successfully completed, insert the DUT (in this example, the
amplifier) into the test setup between the noise source and RF Input of the spectrum
analyzer. (see Fig. 2)
Low pass
Noise Source
DUT
0V
+28V
Noise 28V
RF
Fig. 2 Test Setup