Remote Commands
R&S
®
FPL1000
816
User Manual 1178.3370.02 ─ 03
8.10.1 Configuring the Reference Frequency
SOURce:EXTernal:ROSCillator:EXTernal:FREQuency
...................................................... 816
........................................................................................ 816
[SENSe:]ROSCillator:SOURce:EAUTo?
........................................................................... 816
SOURce:EXTernal:ROSCillator:EXTernal:FREQuency
<Frequency>
This command defines the frequency of the external reference oscillator.
If the external reference oscillator is selected, the reference signal must be connected
to the rear panel of the instrument.
Parameters:
<Frequency>
Range:
1 MHz to 20 MHz
Example:
ROSC:EXT:FREQ 13MHZ
Sets the frequency to 13 MHz.
SOUR:EXT:ROSC:EXT:FREQ 13MHZ
[SENSe:]ROSCillator:SOURce
<Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S
FPL1000.
Parameters:
<Source>
INTernal
The internal reference is used (10 MHz)
EXTernal
The external reference from the REF INPUT 10 MHZ connector
is used; if none is available, an error flag is displayed in the sta-
tus bar
Example:
ROSC:SOUR EXT
[SENSe:]ROSCillator:SOURce:EAUTo?
This command queries the current reference type in case you have activated an auto-
matic switch to the internal reference if the external reference is missing.
Parameters:
<Reference>
INT
internal reference
EXT
external reference
Example:
SENS:ROSC:SOUR:EAUT?
Queries the currently available reference type.
Usage:
Query only
Configuring the R&S
FPL1000