Remote Commands
R&S
®
FPL1000
815
User Manual 1178.3370.02 ─ 03
MMEM:CAT? '*.xml'
//Returns a list of all xml files in the directory 'C:\R_S\INSTR\USER\Results'.
MMEM:CAT:LONG? '*.xml'
//Returns additional information about the xml files in the directory 'C:\R_S\INSTR\USER\Results'.
8.9.6.3
Storing Instrument Settings
In this example we will store the instrument settings for the "Spectrum" channel setup.
INST:SEL 'SPECTRUM'
//Selects measurement channel 'SPECTRUM'.
MEMM:STOR:TYPE CHAN
//Specifies that channel-specific data is to be stored.
MMEM:STOR:STAT 1, 'C:\R_S\Instr\user\Spectrum'
//Stores the channel settings from the 'Spectrum' channel
// to the file 'Spectrum.dfl'.
8.9.6.4
Loading Instrument Settings
In this example we will load the hardware settings from the configuration file
Spectrum.dfl
to a new "Spectrum2" channel setup.
MEMM:LOAD:TYPE NEW
//Specifies that settings will be loaded to a new channel besides the existing
//'Spectrum' channel.
MMEM:SEL:CHAN:HWS ON
//Selects only hardware settings to be loaded.
MMEM:LOAD:STAT 1, 'C:\R_S\Instr\user\Spectrum'
//Loads the channel-specific settings from the file 'C:\R_S\Instr\user\Spectrum.dfl'
//to a new channel. The new channel is named 'Spectrum2' to avoid a naming conflict
//with the existing 'Spectrum' channel.
INST:REN 'Spectrum2','Spectrum3'
//Renames the loaded channel to 'Spectrum3'.
8.10 Configuring the R&S
FPL1000
The remote commands required to set up the R&S
FPL1000 are described here.
Configuring the Reference Frequency
.................................................................. 816
.........................................................................................817
.....................................................................................819
............................................................................ 823
Remote Commands for Language Settings
..........................................................830
Configuring the Network and Remote Control
...................................................... 830
Checking the System Configuration
......................................................................834
....................................................................................... 839
Configuring the R&S
FPL1000