Remote Commands
R&S
®
FPL1000
658
User Manual 1178.3370.02 ─ 03
Parameters:
<Unit>
DBM | V | A | W | DBPW | WATT | DBUV | DBMV | VOLT |
DBUA | AMPere
*RST:
dBm
Example:
UNIT:POW DBM
Sets the power unit to dBm.
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level (for all traces in all windows).
With a reference level offset
≠
0, the value range of the reference level is modified by
the offset.
Suffix:
<n>, <t>
.
irrelevant
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>, <t>
.
irrelevant
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
" Shifting the Display ( Offset )"
8.8.3.2
Configuring the Attenuation
...............................................................................................659
INPut:ATTenuation
<Attenuation>
This command defines the total attenuation for RF input.
Setting Basic Spectrum RF Measurement Parameters