Remote Commands
R&S
®
FPL1000
616
User Manual 1178.3370.02 ─ 03
//Queries the position of the harmonics.
CALC:MARK:FUNC:HARM:DIST? TOT
//Queries the total harmonic distortion.
8.6.10 Measuring the Third Order Intercept Point
..............................................................................................616
Programming Example: Measuring the TOI
..........................................................617
8.6.10.1
Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
●
●
CALCulate<n>:DELTamarker<m>:X:RELative?
●
CALCulate<n>:DELTamarker<m>:Y?
●
●
Remote commands exclusive to TOI measurements
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]
......................................................... 616
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE
...................................... 617
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
......................................................... 617
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]
<State>
This command initiates a measurement to determine the third intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instru-
ment. Marker 1 and marker 2 (both normal markers) are set to the maximum of the two
signals. Delta marker 3 and delta marker 4 are positioned to the intermodulation prod-
ucts. The delta markers can be modified separately afterwards with the
command.
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:MARK:FUNC:TOI ON
Switches on the measurement of the third-order intercept.
Configuring and Performing Measurements