Remote Commands
R&S
®
FPL1000
606
User Manual 1178.3370.02 ─ 03
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
8.6.8.2
Performing a Time Domain Power Measurement
The following commands are required to perform a Time Domain Power measurement:
Chapter 8.6.1, "Performing Measure-
8.6.8.3
Retrieving Measurement Results
The following commands query the results for time domain measurements.
Measuring the Mean Power
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:AVERage:RESult?
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?
.......................... 606
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?
......................................607
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:AVERage:RESult?
This command queries the average mean time domain power. The query is only possi-
ble if averaging has been activated previously using
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<MeanPower>
Mean power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?
This command queries the maximum mean time domain power. The query is only pos-
sible if the peak hold function has been activated previously using
MARKer<m>:FUNCtion:SUMMary:PHOLd
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
Configuring and Performing Measurements