Remote Commands
R&S
®
FPL1000
574
User Manual 1178.3370.02 ─ 03
Suffix:
<sb>
.
1 | 2 | 3
<range>
1..30
Selects the SEM range.
<PClass>
1..4
Power class for which the limit is defined.
Parameters:
<Level>
Relative limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
-50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
Chapter 8.6.6.10, "Example: SEM
Manual operation:
See
" Rel Limit Start / Stop <n> "
[SENSe:]ESPectrum<sb>:RANGe<range>:LIMit<PClass>:RELative:STOP:ABS
<Limit>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
For more information see
"Relative limit line functions"
Suffix:
<sb>
.
1 | 2 | 3
<range>
1..30
Selects the SEM range.
<PClass>
1..4
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the stop frequency of a SEM range to be used
in addition to the relative limit if the MAX function is enabled (see
[SENSe:]ESPectrum<sb>:RANGe<range>:
LIMit<PClass>:RELative:STOP:FUNCtion
on page 575).
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
ABSolute -15
Chapter 8.6.6.10, "Example: SEM
Manual operation:
See
" Rel Limit Start / Stop <n> "
Configuring and Performing Measurements