Remote Commands
R&S
®
FPL1000
560
User Manual 1178.3370.02 ─ 03
//Initiates a new measurement and waits until the sweep has finished.
//---------------Retrieving Results-------------
CALC:MARK:FUNC:POW:RES? CN
//Returns the carrier-to-noise ratio.
8.6.5 Measuring the Occupied Bandwidth
All remote control commands specific to occupied bandwidth measurements are
described here.
...............................................................................560
Programming Example: OBW Measurement
........................................................561
8.6.5.1
Configuring the Measurement
The following commands configure measurements of the occupied bandwidth.
Useful commands for occupied bandwidth measurements described elsewhere
Configuring the channel:
●
[SENSe:]POWer:ACHannel:BANDwidth[:CHANnel<ch>]
●
●
[SENSe:]POWer:ACHannel:PRESet:RLEVel
Defining search limits:
●
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
●
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
●
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
Performing the measurement:
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer:SELect
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer[:STATe]
Retrieving results:
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
Remote commands exclusive to occupied bandwidth measurements:
................................................................................................ 560
...........................................................................................560
[SENSe:]POWer:BWIDth
<Percentage>
[SENSe:]POWer:BANDwidth
<Percentage>
This command selects the percentage of the total power that defines the occupied
bandwidth.
Configuring and Performing Measurements