Remote Commands
R&S
®
FPL1000
556
User Manual 1178.3370.02 ─ 03
●
Remote commands exclusive to channel power measurements
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ
.............................................. 556
.................................................................................. 556
CALCulate<n>:MARKer<m>:FUNCtion:POWer:RESult:PHZ
<State>
This command selects the way the R&S
FPL1000 returns results for power measure-
ments.
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:
.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
ON | 1
Channel power density in dBm/Hz
OFF | 0
Channel power in dBm
*RST:
0
Example:
CALC:MARK:FUNC:POW:RES:PHZ ON
Output of results referred to the channel bandwidth.
Manual operation:
See
"Channel power level and density ( Power Unit )"
[SENSe:]POWer:ACHannel:MODE
<Mode>
This command selects the way the R&S
FPL1000 displays the power of adjacent chan-
nels.
You need at least one adjacent channel for the command to work.
Parameters:
<Mode>
ABSolute
Shows the absolute power of all channels
RELative
Shows the power of adjacent and alternate channels in relation
to the transmission channel
*RST:
RELative
Manual operation:
See
" Absolute and Relative Values (ACLR Mode) "
Configuring and Performing Measurements