Remote Commands
R&S
®
FPL1000
547
User Manual 1178.3370.02 ─ 03
If you set the channel spacing for a transmission channel, the R&S
FPL1000 sets the
spacing of the lower transmission channels to the same value, but not the other way
round. The command works hierarchically: to set a distance between the 2nd and 3rd
and 3rd and 4th channel, you have to set the spacing between the 2nd and 3rd chan-
nel first.
Suffix:
<ch>
.
1 to 18
Tx channel number
Parameters:
<Spacing>
Range:
14 kHz to 2000 MHz
*RST:
20 kHz
Manual operation:
See
[SENSe:]POWer:ACHannel:TXCHannel:COUNt
<Number>
This command defines the number of transmission channels.
The command works for measurements in the frequency domain.
Parameters:
<Number>
Range:
1 to 18
*RST:
1
Manual operation:
See
" Number of channels: Tx , Adj "
8.6.3.3
Defining Weighting Filters
The following commands define weighting filters for ACLR measurements.
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
....................................................... 547
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL]
...............................................................548
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
................................................548
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<ch>
................................................. 548
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ACHannel
......................................................548
[SENSe:]POWer:ACHannel:FILTer[:STATe][:ALL]
............................................................. 549
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
.............................................. 549
[SENSe:]POWer:ACHannel:FILTer[:STATe]:CHANnel<ch>
................................................549
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
<Alpha>
This command defines the roll-off factor for the adjacent channel weighting filter.
Parameters:
<Alpha>
Roll-off factor
Range:
0 to 1
*RST:
0.22
Manual operation:
See
Configuring and Performing Measurements