Remote Commands
R&S
®
FPL1000
531
User Manual 1178.3370.02 ─ 03
INITiate<n>:SEQuencer:MODE
<Mode>
This command selects the way the R&S
FPL1000 application performs measurements
sequentially.
Before this command can be executed, the Sequencer must be activated (see
Note:
In order to synchronize to the end of a sequential measurement using *OPC,
*OPC? or *WAI you must use
SINGle
Sequence mode.
Suffix:
<n>
.
irrelevant
Parameters:
<Mode>
SINGle
Each measurement is performed once (regardless of the chan-
nel's sweep mode), considering each channels' sweep count,
until all measurements in all active channels have been per-
formed.
CONTinuous
The measurements in each active channel are performed one
after the other, repeatedly (regardless of the channel's sweep
mode), in the same order, until the Sequencer is stopped.
CDEFined
First, a single sequence is performed. Then, only those channels
in continuous sweep mode (
INIT:CONT ON
) are repeated.
*RST:
CONTinuous
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Manual operation:
See
SYSTem:SEQuencer
<State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (
INIT:SEQ...
) are executed, otherwise an error will
occur.