Remote Control
R&S
®
FPC
321
User Manual 1178.4130.02 ─ 07
●
Impedance:
INPut:IMPedance
INPut:IMPedance:PAD
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:PDIVision
.................................................... 321
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RPOSition
................................................... 321
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:PDIVision
<Scale>
This command defines the range of the y-axis for the graphical results.
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Scale>
<numeric value> (integer only)
The unit depends on the result.
•
PCT for ASK results (1 % to 100 % per grid division).
•
Hz for FSK results (1 kHz to 1 MHz per grid division).
Example:
//Define grid scale of AM results (50 percent per grid)
CALC:FEED 'XTIM:DDEM:MEAS'
DISP:TRAC:Y:PDIV 50
Manual operation:
See
DISPlay<n>[:WINDow]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the position of the reference level on the display grid.
Prerequisites for this command
●
Select ASK demodulation (
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Position>
<numeric value> (integer)
Range:
1 to 100
*RST:
100
Default unit: PCT
Example:
//Define position of reference level
DDEM ASK
CALC:FEED 'XTIM:DDEM:MEAS'
DISP:TRAC:Y:RPOS 80
Manual operation:
See
Remote Commands of the Digital Demodulation